LT Syringe Pump

We put the best materials and designs together using manufacturing processes to ensure optimum alignment of the drive, seal and syringe barrel. The end result is a high value syringe pump that is “elegant in its simplicity.” The LT syringe pump has all the accuracy and precision characteristics of ordinary syringe pumps without syringe replacement schedules and user accessibility concerns. You enjoy longer instrument “up time” and the freedom to design instruments without consuming valuable working area.

Every aspect of the LT syringe pump has been specifically designed to assure trouble free operation for the life of the syringe pump.



  • NO calibration required
  • NO high fluid pulsation

  • NO leaking bellows or diaphragms

  • NO ceramics that seize when dry

  • NO consumable syringes



Along with a full range of syringe sizes, there are valve, encoder and electronic options to meet almost every instrument application and budget.

SKU
LT_Syringe_Pump

Specifications

Volume Displacement
(nominal)

250 μL
Catalog # 0386

500 μL 
Catalog # 0387

1 mL 
Catalog # 0388

2.5 mL
Catalog # 0389

5 mL
Catalog # 0390

Step Resolution

0.150 μL / full step 0.300 μL / full step 0.600 μL / full step 1.500 μL / full step 3.000 μL / full step

Syringe Data 
  Bore
  Stroke Fluid
  Backpressure
  Interface



0.1721 in. [4.37 mm]
0.667 in. [16.93 mm]
30 psi max. [2.07 bar]
1/4-28 female port



0.2433 in. [6.18 mm]
0.667 in. [16.93 mm]
30 psi max. [2.07 bar]
1/4-28 female port



0.3446in. [8.75 mm]
0.667 in. [16.93 mm]
20 psi max. [1.38 bar]
1/4-28 female port



0.5441 in. [13.82 mm]
0.667 in. [16.93 mm]
10 psi max. [.689 bar]
1/4-28 female port



0.7694 in. [19.54 mm]
0.667 in. [16.93 mm]
5 psi max. [.344 bar]
1/4-28 female port

Materials 

PPS, Borosilicate Glass, UHMWPE, Viton®




Functional Parameters

Imprecision (full stroke)

< 0.05% CV

Innaccuracy (full stroke)

< 1.0%

Dimensions (max.)
  Overall Width
  Overall Height
  Overall Depth



2.12 in. [53.85 mm]
5.90 in. [149.86 mm]
2.62 in. [66.55 mm]

Motor/Leadscrew Data 
Travel/Step
Step Angle
Resistance/Phase
Inductance/Phase
Rated Current
Rated Voltage 



.0004 in. [0.01 mm] per full step
1.8 deg.
3.5  Ω 
4.0 mH
900 mA
3.2 V*

Limit Sensor Data
  Type
  Diode Current


Optoschmitt (Honeywell HOA6990)
15 mA typical

* Pump should typically be driven with a 24 VDC bipolar chopper drive, ½ step, at rated current. Higher voltage may be required at higher step rates.

LT-Series Pump Interface Details

Dimensions/Features noted are typical for any size, LT-Series Pump (5.0 mL shown)

LT-Series 5.0mL

Please contact your TriContinent Sales Representative for current pricing.

LT Syringe Pump with No Electronics and No Valve

Catalog #Description
0386 LT Pump, 250 μL
0387 LT Pump, 500 μL
0388 LT Pump, 1.0 mL
0389 LT Pump, 2.5 mL
0390 LT Pump, 5.0 mL

M-Series Controller PCBA

Catalog #Description
0956 Controller/Driver Assembly (.5A Max)
0959 RS232/485 Converter Kit
0960 USB/485 Converter Kit


 

LT-Series Warranty Information

TriContinent operates an ISO 13485 registered manufacturing facility that delivers the highest quality products.  LT-Series Pumps are general laboratory components not subject to FDA regulatory approval.  TriContinent uses the highest quality components with applicable safety markings to meet the needs of system designers.  LT-Series Pumps do not carry independent CE markings as they are components designed for use within larger systems which are subject to independent testing and certification.  

We stand behind the LT-Series Pumps with a 1-year warranty.  During the warranty period, should a LT-Series Pump fail, call us with an explanation of malfunction and TriContinent will issue a Return Material Authorization (RMA) number for repair or replacement at our discretion, at no charge.

It is the responsibility of the customer to determine the suitability of an application and the material compatibility of our product with your application.

When returning a product for repair or replacement, please contact TriContinent Customer Service to receive an RMA number.  When calling please be sure to have the following information available:

          Catalog Number
         Serial Number
         Explanation of Malfunction

TriContinent will not accept the return of any product with out prior approval. TriContinent reserves the right to refuse a return of product that has been used with infectious microbiological or radioactive substances or any other materials that may be deemed hazardous.

The RMA number should be referenced on the outside of the shipping container when returning to TriContinent.

 

Warranty Exclusions

The above warranty shall not apply for defective products resulting from:

         Improper or Inadequate maintenance by the customer
         Unauthorized maintenance or service
         Customer misuse
         Operation outside of the environmental specifications for the product
         Instruments returned without proper packaging

There is no warranty expressed for the following components:

         Syringes
         Plungers

Note: For equipment damaged in shipping, please contact the shipper directly.

Thank you for your interest in Gardner Denver. Please fill out the form below and we will get back to you as soon as possible.

Your request will be forwarded to the relevant individuals to assist with your request.

By clicking 'Send' you are authorizing Gardner Denver to contact you so that we may fulfill your request, as well as to communicate with you regarding our products and services.

Terms & Conditions

Privacy Policy